Abstract
Background: Measurable residual disease (MRD) analysis is critical in acute lymphoblastic leukaemia (ALL) management protocols. Real-time quantitative polymerase chain reaction (RT-qPCR) is the accepted gold standard in MRD testing. Multi-parameter flow cytometry has emerged as a comparative testing modality. We evaluated the concordance of our centre’s 10-colour flow cytometry (Navios and Navios EX) with RT-qPCR (ABI 7500 Fast Real-Time PCR) for MRD detection.
Aim: To investigate the concordance of MRD detection using flow cytometry and RT-qPCR; and, separately, to compare immunoglobulin heavy chain (IgH) rearrangement PCR with RT-qPCR.
Setting: The National Health Laboratory Services Flow Cytometry and Somatic Cell Genetics Units at Charlotte Maxeke Johannesburg Academic Hospital.
Methods: Flow cytometry results between March 2020 and June 2023 were compared with RT-qPCR results in follow-up samples of patients with a diagnosis of B-cell acute lymphoblastic leukaemia (B-ALL), with one of the following genetic lesions: t(1;19), t(12;21) or t(9;22).
Results: A total of 74 bone marrow aspirate samples were eligible for inclusion. The flow cytometry assay showed a low sensitivity (13%) but a high specificity (98%) when compared with RT-qPCR. The positive and negative predictive values were 75% and 71% for flow cytometry versus qPCR, below the acceptable levels of 80% and 90%, respectively. Compared with RT-qPCR, the immunoglobulin heavy (IgH) assay demonstrated higher sensitivity (43%) and positive predictive value (91%), although sensitivity remained below levels for clinical utility.
Conclusion: The flow cytometry assay at our centre exhibited significantly lower sensitivity compared to PCR-based methods for the detection of MRD in follow-up cases. Among the PCR techniques, RT-qPCR demonstrated superior sensitivity for MRD detection in B-ALL.
Contribution: The research demonstrates the limitations of the current flow cytometry assay compared with RT-qPCR and identifies the need for improved MRD-tailored approaches, particularly in resource-limited settings where molecular testing may not be readily available.
Keywords: B-ALL; flow cytometry; measurable residual disease; real-time quantitative polymerase chain reaction; immunoglobulin heavy rearrangement.
Introduction
Acute lymphoblastic leukaemia (ALL) is the most common type of cancer in children. It accounts for about 25% of cancers in children under the age of 15 years1 and less than 1% of all adult cancers.2 A critical part of the management of ALL includes the assessment of measurable residual disease (MRD) at several time points during the treatment protocol. Measurable residual disease is recognised as an independent predictor of survival in ALL3,4 and is regarded as the most reliable predictor of relapse-free survival.5 In ALL, early negativity for MRD is associated with very good outcomes in both children and adults.6 In essence, MRD provides useful information for prognostication, guiding therapy and risk stratification of patients.7
Advancements in technology and understanding of the underlying disease pathogenesis have led to the development of different techniques for the detection of MRD. Examples include various polymerase chain reaction (PCR) based techniques such as digital droplet PCR, real-time quantitative PCR (RT-qPCR), next-generation sequencing (NGS) and flow cytometry-based methods. The sensitivity and specificity vary among the different MRD testing modalities, with RT-qPCR recognised as the gold standard for the detection of MRD in MRD-based ALL protocols.8,9 Several studies evaluated the correlation of MRD in B-ALL and outcome at different treatment time points and MRD levels (> 1%, 0.01% – 1% and < 0.01%) and validated an MRD threshold of 0.01 (10−4) to be clinically significant, namely to identify patients who are at a higher risk of relapse.10
Multi-parameter flow cytometry was initially established as a method for MRD assessment, with subsequent studies demonstrating good correlation between flow cytometry findings and RT-qPCR MRD results.11,12,13 Flow cytometry has several advantages over RT-qPCR, namely a shorter turnaround time, less labour-intensive and wider availability.3 Thus, in resource-limited areas, particularly in Africa and other developing countries, multi-parameter flow cytometry is an appealing modality for MRD monitoring where RT-qPCR is not readily available.
In this study, we evaluated the sensitivity and specificity of the Beckman Coulter ClearLab 10-colour flow cytometry assays performed on the Beckman Coulter Navios and Navios EX flow cytometry system for MRD monitoring or detection in comparison with the gold standard RT-qPCR. Additionally, we evaluated the concordance between RT-qPCR and immunoglobulin heavy (IgH) PCR MRD testing.
Research methods and design
Study setting
The National Health Laboratory Services (NHLS) operates specialised diagnostic units – including the flow cytometry laboratory and Somatic Cell Genetics Units (SCGU) – at the Charlotte Maxeke Johannesburg Academic Hospital (CMJAH) in Johannesburg, South Africa. These laboratories are South African National Accreditation System (SANAS) (International Organization for Standardization [ISO]) 15189:2022 accredited with an associated subscription to both local and international external quality assurance (EQA) testing programmes.
Study design and study population
A retrospective study was conducted to review and analyse the degree of concordance between flow cytometry and RT-qPCR, as well as between flow cytometry and IgH PCR, in B-ALL MRD follow-up cases performed on bone marrow aspirate samples from patients in morphological remission (< 5% blasts). The study included paediatric and adult patients with a diagnosis of B-ALL, bearing one of the following recurrent genetic lesions; the translocation t(1;19)(q23;p13.3); TCF3::PBX1 (t[1;19]), translocation t(12;21)(p13.2;q22.1); ETV6::RUNX1 (t[12;21]) or translocation t(9;22)(q34.1;q11.2); BCR::ABL1 (t[9;22]), for whom flow cytometry and RT-qPCR were performed at follow-up for MRD testing at the same time point, at the CMJAH NHLS laboratories. The study period was from March 2020 to June 2023. Given that MRD monitoring occurs at several different time points during therapy (i.e. post-induction, post-consolidation, post-maintenance), each time point was viewed separately for the comparisons between the different platforms.
The patient samples were analysed on the Navios and Navios EX (Beckman Coulter, San Diego, CA, United States [US]) flow cytometer, using the Kaluza C™ software platform (Kaluza Ltd, London, United Kingdom) at the CMJAH NHLS flow cytometry laboratory. The RT-qPCR was performed on the ABI 7500 Fast Real-Time PCR system (Applied BiosystemsTM, Waltham, MA, US) at the SCGU. A list of RT-qPCR tests performed during the specified time frame was obtained from the SCGU database. Each item in the list was screened for eligibility alongside the data obtained from the NHLS laboratory information system. The eligible patient entries and data were transcribed onto an Excel spreadsheet. Additional flow cytometry data (number of events analysed for each case and panel[s] used) were obtained from analysis files generated by the Navios and Navios EX instruments via the Kaluza C™ software interface.
The backbone B panel used for all the cases contained cluster of differentiation (CD) markers: CD5, CD10, CD19, CD20, CD34, CD38, CD45, CD200, kappa and lambda. The target median (interquartile range [IQR]) of events analysed (excluding nucleated red blood cells) was 50 000 events. Other panels commonly used in addition to the backbone B panel, included the M2 tube, which contained the following antibodies: CD13, CD15, CD19, CD33, CD34, CD45, CD38, CD117, CD123 and HLA-DR; and the Haematogone (H) panel containing the following markers: CD10, CD19, CD22, CD34, CD45, CD58 and CD81.
Real-time quantitative polymerase chain reaction was performed on the ABI 7500 Fast Real-Time PCR system using fluorescent-based chemistry for the qualitative and quantitative detection of fusion transcripts. RT-qPCR generated complementary deoxyribonucleic acid (cDNA) products from ribonucleic acid (RNA) transcripts using Invitrogen SuperScript IV VILO master mix with ezDNase (Thermo Fisher Scientific, Waltham, MA, US), with quantitation based on fluorescence detection during amplification, using amplification end points and dissociation curves for analysis. The results were analysed with percentages and ratios of Fusion gene to control gene. The control gene, GUSB, was used to gauge sample degradation (defined as < 1000 copies of GUSB), validate any undetectable fusion transcript result and calculate fusion transcript ratios.14
The IgH assay used was a PCR-based assay employing the IdentiClone® IgH Gene Clonality Assay (Invivoscribe, Inc., San Diego, CA, US), which contains multiple consensus DNA primers that target the variability-diversity-joining (VDJ) segments of the IgH gene to detect clonal rearrangements.15
The following samples were excluded: morphologically haemodiluted samples where the flow cytometry and RT-qPCR failed to detect MRD, samples for which RT-qPCR failed (low RNA yield defined as < 1 ug of RNA),14 cases with morphologic evidence of residual disease and cases without a baseline positive RT-qPCR result. An exception was made for five cases where the blast percentage was more than 5% morphologically, ranging from 6% to 14%. In all cases, there was no detectable disease on flow cytometry; thus, comparison could be made with the corresponding RT-qPCR result to distinguish MRD from normal haematogones or recovering myeloid blasts.
Two separate statistical analyses were performed, the first comparing flow cytometry with the gold-standard RT-qPCR, and the second comparing IgH rearrangement studies with RT-qPCR. No fixed cut-off was applied when comparing flow cytometry with RT-qPCR, as the number of acquired flow cytometry events varied between patients, resulting in variable analytical sensitivity. Measurable residual disease status was therefore assessed by direct comparison with the corresponding RT-qPCR result. Real-time-quantitative polymerase chain reaction analysis was performed in triplicate using a relative quantitation approach. The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) for each analysis were calculated and presented in two separate ‘two-way-tables’. Descriptive statistics (median, IQR, minimum and maximum values) were used to summarise results.
Ethical considerations
Ethics clearance for this study was obtained from the University of the Witwatersrand Human Research Ethics Committee under clearance certificate number M230840. The data were de-identified, with patient identifying information stored in a separate password-protected Excel spreadsheet, which was only accessible to the investigators.
Results
Of the 303 bone marrow aspirate samples, a total of 74 cases in which both flow cytometry and RT-qPCR results were available were eligible for inclusion in the study (Figure 1). Two hundred and twenty-nine samples were excluded due to: (1) absence of a corresponding flow cytometry analysis performed at the CMJAH NHLS flow cytometry laboratory (n = 185), (2) sample degradation (n = 9), (3) insufficient RNA yield (n = 19), (4) lack of a baseline positive RT-qPCR result (n = 13) or (5) haemodilution (n = 3). The median age (IQR) of the cohort was 5 years (2–9 years); all the samples were from children and adolescents except for one adult patient at age 28 years. There were an almost equivalent number of male and female patients. The most common genetic aberration was t(12;21) constituting 52.6% (n = 40) of the patients, followed by t(1;19) at 44.6% (n = 33) and t(9;22) at 2.6% (n = 1). Corresponding RT-qPCR assays specific for each fusion transcript (ETV6::RUNX1, TCF3::PBX1 and BCR::ABL1) were performed for the respective cases. As paediatric patients from the referring centres are treated with the ALL Berlin–Frankfurt–Münster (BFM) 95 protocol, without the use of immunotherapeutic targets, this precluded the need for modifications or special considerations of the flow cytometry analysis. In the majority of the samples, a corresponding IgH gene rearrangement PCR had been performed (n = 70); of these, two failed amplifications, leaving 68 results (97%) available for comparative analysis with RT-qPCR and flow cytometry.
 |
FIGURE 1: Bone marrow sample inclusion and exclusion flow chart. |
|
Flow cytometry versus real-time quantitative polymerase chain reaction
A total of 74 cases were assessed for comparison of flow cytometry versus RT-qPCR. The majority of cases (~81%) tested for MRD by flow cytometry utilised two separate antibody panels, including a premixed ClearLLab 10-colour backbone B panel in addition to the M2 tube and Haematogone panel. The use of additional panels during flow cytometry was dependent on the tumour-specific immunophenotype, which was reported at presentation. The M2 tube was included for aberrant myeloid markers (if present) or to assess for CD123 and HLA-DR expression, which are frequently expressed by leukaemic cells.
The median (IQR) of events analysed (excluding nucleated red blood cells) was 50 000 events.
Of the 74 cases included, 23 were MRD positive, and 51 were MRD negative by RT-qPCR. The flow cytometry assay showed low sensitivity (13%) but a high specificity (98%) as compared with RT-qPCR (Table 1). The PPV and NPV values were 75% and 71%, below the acceptable levels of 80% and 90%, respectively. The RT-qPCR results of the false-negative cases were below the limit of detection (LOD) of the flow cytometry assay (0.04%), with a median (IQR) ratio of Fusion gene:GUSB of 0.55 × 10−3 (0.08 to 2.60 × 10−3). Application of a clinically significant RT-qPCR threshold of 0.01% increased the sensitivity of flow cytometry to 60%, while the specificity remained high at 99%, with PPV and NPV values of 75% and 97%, respectively.
| TABLE 1: Two-way table to calculate sensitivity, specificity, positive and negative predictive values for Navios flow cytometry versus real-time quantitative polymerase chain reaction. |
For the three true positive cases (two cases with t(12;21) and one case with t(1;19) – see Figure 2, Table 1-A1 Appendix 1), the amount of disease detected on flow cytometry was between 0.2 and 1% of total events analysed, while the median (IQR) ratio of Fusion gene:GUSB was 0.014 (0.0095 to 0.027), which is higher than the Fusion gene:GUSB ratio in the false-negative cases (Table 1-A2). Six samples were reported as haemodilute, and all were negative for MRD on flow cytometry. Half of these samples (n = 3) had detectable MRD on RT-qPCR, and thus the latter were included in the overall analysis. The Cohen’s kappa coefficient calculated to assess agreement between flow cytometry testing and RT-QPC showed poor agreement, K = 0.10 (95% confidence interval [CI]: 0.08987 to 0.29079, s.e. (standard error): 0.09711).16
 |
FIGURE 2: Truth analysis – Flow cytometry-based measurable residual disease vs real-time quantitative polymerase chain reaction. |
|
A small subset (n = 5) of the samples had no immunophenotypic evidence of disease; however, morphologically there were more than 5% blasts. For three out of the five cases, this finding was accounted for as normal reactive precursor B-cells on flow cytometric analysis. This correlated with the corresponding RT-qPCR and IgH results, which were also negative for MRD.
Real-time quantitative polymerase chain reaction versus immunoglobulin heavy
The IgH assay showed low sensitivity (43%) as compared with RT-qPCR but higher than flow cytometry, and 98% specificity (Table 2 and Figure 3). The PPV was 91%, but the NPV was suboptimal at 77%. The RT-qPCR results of the false-negative cases were below the LOD of the IgH assay (0.01%) with a median (IQR) ratio of Fusion gene:GUSB of 0.3 × 10−3 (0.025 to 0.7 × 10−3). In contrast, in the true positive cases, the median (IQR) of the ratio of Fusion gene:GUSB was expectedly higher at 3.0 × 10−3 (0.45–13.5 × 10−3), with 0.004 the lowest value obtained where both RT-qPCR and IgH were positive. The calculated Cohen’s kappa coefficient was 0.47 (95% CI: 0.25617–0.68957, s.e.: 0.11056), which denotes a moderate strength of agreement between IgH PCR and RT-qPCR.16
 |
FIGURE 3: Truth analysis – Immunoglobulin heavy chain rearrangement polymerase chain reaction – Based on measurable residual disease vs real-time quantitative polymerase chain reaction. |
|
| TABLE 2: Two-way table to calculate sensitivity, specificity, positive and negative predictive values for immunoglobulin heavy chain rearrangement polymerase chain reaction versus real-time-quantitative polymerase chain reaction. |
Discussion
Monitoring for MRD is pivotal in the management of ALL and is the standard of care in ALL treatment protocols. Results of MRD testing alter risk, prognostication and treatment approaches. As these data become available, clinical decisions are made, including identifying patients at high risk of disease relapse and those who are eligible for allogeneic stem cell transplantation.
The recommended sensitivity levels for MRD testing range from 10−4 to 10−5, corresponding to the detection of 1 leukaemic cell per 10 000 and 100 000 cells, respectively.17 Historically, RT-qPCR techniques have met the desired sensitivity levels, and with the advent of improved high-throughput multi-parameter flow cytometry assays, several reports have shown correlation in the sensitivities between molecular modalities and flow cytometry.3,5,18
As our flow cytometry assay is not MRD tailored, the sensitivity results are not unexpected, with only around 13% of MRD-positive cases detected. These cases had a higher fusion transcript burden and fusion transcript-to-GUSB ratio compared with the false-negative cases, reflecting the low sensitivity of the assay.
According to the Australasian Cytometry Society Guidelines, for a lower LOD, the minimum number of total events needed for a sensitivity level of 0.01% (10−4) is 2 × 105 cells, while the minimum number of leukaemic targets or tumour events required for an MRD-positive result is 20. For the lower limit of quantification (LOQ), the minimum number of total events needed for a sensitivity level of 0.01% (10−4) is 2 × 105 cells, while the minimum number of target cells required for an MRD-positive result is 50.19 Most MRD protocols are based on a sensitivity of 0.01% (10−4) to 0.001% (10−5).17,20 If the recommendations are applied to our flow cytometry assay, where a minimum of 50 000 events are analysed, the LOD and LOQ would be below the desirable sensitivity at 0.04% and 0.1%, respectively. To meet the sensitivity requirements as an MRD assay, the number of CD45-positiveevents would need to increase to 500 000, 10 times more than the current target.
Our flow cytometry assay showed high specificity of close to 100% as compared with the gold standard, demonstrating the low probability of calling MRD positivity when not present, that is, low false-positive rates. Ultimately, molecular assays would need to be awaited in flow cytometry-negative cases to guide clinical decision-making in our setting. In similar settings with resource limitations, flow cytometry assays, while still not meeting current guidelines for MRD testing, may provide some benefit as an interim mode of assessment while results of samples referred for molecular tests are awaited. Given that flow cytometry has shorter turnaround times, is less labour-intensive and is more widely available than RT-qPCR, it remains an attractive modality for MRD assessment, particularly in cases lacking a recurrent genetic abnormality. In addition, the comparatively greater volume of research, broader standardisation efforts and more established external quality assessment (EQA) programmes related to MRD flow cytometry support an anticipated shift towards greater use of flow cytometry for MRD assessment.20
The PPV and NPV for flow cytometry were suboptimal, with the clinical utility limited to cases where MRD is detected. The single false-positive case (flow cytometry versus RT-qPCR) showed a 5% population of cells on flow cytometry with a similar immunophenotype to that seen at presentation and was therefore concluded to represent residual disease. Both the IgH PCR and RT-qPCR techniques were negative for MRD. In this specific case, the difference from normal immunophenotype was CD38 under-expression and CD200 over-expression but with no non-B-cell lineage aberrancies. This highlights the challenge in delineating disease from normal B-cell precursors morphologically in the bone marrow and immunophenotypically on flow cytometry, the need for caution in interpreting MRD on flow cytometry when no definite aberrancies are present and the requirement to confirm suspected MRD with an alternative method such as RT-qPCR in these cases. For this case, both PCR techniques were negative for MRD. This may indicate that while assessment of cell population scatter patterns and associated aberrancies is valuable in delineating disease from normal B-cell precursors in the bone marrow, it is not definitive or absolute. This is particularly true, as clonal evolution is observed post-chemotherapy with antigen expression changes that can affect MRD testing.4,7 This highlights the important role PCR techniques play in further clarifying such cases.
For the discordant flow cytometry and RT-qPCR result, there were 5% blasts on morphology; the RT-qPCR was positive for MRD, while the flow cytometry did not detect disease. However, a small abnormal population of blasts (CD34 positive with asynchronous expression of CD20) was detected, reflecting a shift in antigen expression in the disease population. Several studies have examined these shifts in detail to optimise their panels, a practice that ensures MRD with antigen shift is detected.3 The use of novel immunotherapeutic agents such as anti-CD19, anti-CD22 therapy and chimeric antigen receptor (CAR) T-cell therapy further poses challenges to MRD monitoring due to key antigen (particularly CD19) loss. This requires specially tailored panels and algorithms that will be able to carefully discriminate disease populations, as shown in one study that employed a special all-colour panel consisting of 10 antibodies (namely CD10, CD19, CD20, CD22, CD24, CD34, CD38, CD45, CD58 and cytoplasmic CD79a) and Syto41, a DNA dye to identify nucleated cells. The combination of CD79a and CD22 was used as a main gating substitute for CD19, while CD10 and CD24 were alternatives if the tumour population expressed none of these antigens.7
As expected, IgH rearrangement testing, being a PCR technique with a higher analytical sensitivity than flow cytometry due to its ability to detect very low-level disease beyond the event limitations of the flow cytometric platform, demonstrated superior test performance but with comparable specificity. In contrast to flow cytometry, the PPV fared better than the NPV, with no false-positive cases. Thus, a positive result on IgH is reliable, while a negative result does not exclude MRD, and alternative, more sensitive molecular techniques are required in the latter.
The reliability of MRD testing results is heavily dependent on the quality of samples analysed. Haemodilute samples are typically contaminated with peripheral blood and may yield falsely negative results. From our results, it would also appear that more sensitive MRD testing modalities such as RT-qPCR are more likely to detect disease in the face of sample haemodilution. This is also relevant and useful for post-induction bone marrow samples, which are usually pauci-cellular and have a limited number of events on flow cytometry. Haemodilution was determined on morphology alone, and it is important to note that the quality of each sample submitted for the different testing modalities may not be uniform. It is therefore recommended that MRD reports should contain a comment on the degree of haemodilution, when this assessment is possible.21 This also speaks to the pre-analytical aspects that should be considered when samples are collected, so that optimal and representative samples are obtained. The Australasian Cytometry Society recommends the first-pull bone marrow aspirate for MRD testing.19 In our context, preference should be given to first-pull samples, while IgH or flow cytometry should be prioritised in cases lacking a recurrent genetic abnormality (detectable by RT-qPCR). Therefore, MRD testing should not only focus on the analytical process but should be seen as a comprehensive workflow that seeks to streamline all the pre-analytical, analytical and post-analytical processes to ensure accurate and sensitive MRD results for clinical decision-making.
Limitations
Limitations of this study include:
- Patient files were not examined; thus, evaluation of clinical outcomes in relation to laboratory results was not assessed.
- The number of events analysed may vary between the cases. The maximum number of events analysed was set at 50 000 on the Navios and Navios EX instruments.
Conclusion
The flow cytometry assay employed at our centre exhibited significantly lower sensitivity compared with PCR-based methods for the detection of MRD in follow-up cases of ALL. Among the PCR techniques, RT-qPCR demonstrated superior sensitivity for MRD detection. There is a need to modify the current flow cytometry assay to be tailored specifically for MRD testing by adapting it to accepted standards or guidelines. This would be of value as it would offer alternative methods of MRD testing that are comparatively sensitive for cases not amenable to RT-qPCR testing. As modification of the assay is underway, future studies could conduct a similar comparison looking at sensitivity or specificity of the modified MRD-tailored flow cytometry assay versus RT-qPCR or other PCR techniques.
Acknowledgements
This article is based on research originally conducted as part of Tsholofelo L Malaka’s, master’s research article titled ‘Evaluation of diagnostic sensitivity and specificity of 10-colour flow cytometry testing for disease monitoring/detection on the Navios/Navios EX, Kaluza C platform in comparison to real-time quantitative polymerase chain reaction testing’, submitted to the Faculty of Health Sciences, Department of Molecular Medicine and Haematology, University of the Witwatersrand in 2025. The thesis is currently unpublished and not publicly available. The research article was supervised by Kate Hodkinson and Anima Baiden. The research article was reworked, revised and adapted into a journal article for publication.
Competing interests
The authors declare that they have no financial or personal relationships that may have inappropriately influenced them in writing this article.
CRediT authorship contribution
Tsholofelo L. Malaka: Data curation, Formal analysis, Writing – original draft. Anima Baiden: Conceptualisation, Methodology, Supervision, Writing – review & editing. Katherine E. Hodkinson: Conceptualisation, Methodology, Supervision, Writing – review & editing. All authors reviewed the article, contributed to the discussion of results, approved the final version for submission and publication and take responsibility for the integrity of its findings.
Funding information
This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors.
Data availability
The data that support the findings of this study are available on request from the corresponding author, Tsholofelo L. Malaka.
Disclaimer
The views and opinions expressed in this article are those of the authors and are the product of professional research. It does not necessarily reflect the official policy or position of any affiliated institution, funder, agency or that of the publisher. The authors are responsible for this article’s results, findings, and content.
References
- Kakaje A, Lhalabi MM, Ghareeb A, et al. Rates and trends of childhood acute lymphoblastic leukaemia: An epidemiology study. Sci Rep. 2020;10:6756. https://doi.org/10.1038/s41598-020-63528-0
- Bassan R, Gatta G, Tondini C, Willemze R. Adult acute lymphoblastic leukaemia. Crit Rev Oncol Hematol. 2004;50(3):223–261. https://doi.org/10.1016/j.critrevonc.2003.11.003
- Correia RP, Bento LC, De Sousa FA, Barroso RS, Campregher PV, Bacal NS. How I investigate minimal residual disease in acute lymphoblastic leukemia. Int J Lab Hematol. 2021;43(3):354–363. https://doi.org/10.1111/ijlh.13463
- Modvig S, Hallböök H, Madsen HO, et al. Value of flow cytometry for MRD-based relapse prediction in B-cell precursor ALL in a multicenter setting. Leukemia. 2021;35(7):1894–1906. https://doi.org/10.1038/s41375-020-01100-5
- Tembhare PR, Subramanian PG, Ghogale S, et al. A high-sensitivity 10-color flow cytometric minimal residual disease assay in B-lymphoblastic leukemia/lymphoma can easily achieve the sensitivity of 2-in-106 and is superior to standard minimal residual disease assay: A study of 622 patients. Cytometry B Clin Cytom. 2020;98(1):57–67. https://doi.org/10.1002/cyto.b.21831
- Della Starza I, Chiaretti S, De Propris MS, et al. Minimal residual disease in acute lymphoblastic leukemia: Technical and clinical advances. Front Oncol. 2019;9:726. https://doi.org/10.3389/fonc.2019.00726
- Mikhailova E, Illarionova O, Komkov A, et al. Reliable flow-cytometric approach for minimal residual disease monitoring in patients with B-cell precursor acute lymphoblastic leukemia after CD19-targeted therapy. Cancers (Basel). 2022;14(21):5445. https://doi.org/10.3390/cancers14215445
- Della Starza I, Nunes V, Lovisa F, et al. Droplet digital PCR improves IG-/TR-based MRD risk definition in childhood B-cell precursor acute lymphoblastic leukemia. Hemasphere. 2021;5(3):e543. https://doi.org/10.1097/HS9.0000000000000543
- Rocha JMC, Xavier SG, Souza MEL, Murao M, De Oliveira BM. Comparison between flow cytometry and standard PCR in the evaluation of MRD in children with acute lymphoblastic leukemia treated with the GBTLI LLA – 2009 protocol. Pediatr Hematol Oncol. 2019;36(5):287–301. https://doi.org/10.1080/08880018.2019.1636168
- Contreras YGP, Ostrow TH, Jasinski S, Raetz EA, Carroll WL, Evensen NA. Minimal residual disease in acute lymphoblastic leukemia: Current practice and future directions. Cancers (Basel). 2021;13(8):1847. https://doi.org/10.3390/cancers13081847
- Karawajew L, Dworzak M, Ratei R, et al. Minimal residual disease analysis by eight-color flow cytometry in relapsed childhood acute lymphoblastic leukemia. Haematologica. 2015;100(7):935–944. https://doi.org/10.3324/haematol.2014.116707
- Theunissen P, Mejstrikova E, Sedek L, et al. Standardized flow cytometry for highly sensitive MRD measurements in B-cell acute lymphoblastic leukemia. Blood. 2017;129(3):347–357. https://doi.org/10.1182/blood-2016-07-726307
- Huang Y, Coustan-Smith E, Kao H, et al. Concordance of two approaches in monitoring of minimal residual disease in B-precursor acute lymphoblastic leukemia: Fusion transcripts and leukemia-associated immunophenotypes. J Formos Med Assoc. 2017;11(10):774–781. https://doi.org/10.1016/j.jfma.2016.12.002
- Van Zijl H. Procedure for performing a quantitative real-time PCR on the ABI 7500 Fast Real-Time PCR System to quantify the levels of fusion gene transcripts in haematological malignancies. Johannesburg: National Health Laboratory Service; 2023. Internal SOP.
- Invivoscribe, Inc. IdentiClone® IGH gene clonality assay – ABI fluorescence detection [homepage on the Internet]. No date [cited 2025 Mar 30]. Available from: https://catalog.invivoscribe.com/product/identiclone-igh-gene-clonality-assay-abi-fluorescence-detection/
- MedCalc Software Ltd. Inter-rate agreement [homepage on the Internet]. Version 23.3.7. [cited 2026 May 30]. Available from: https://www.medcalc.org/en/calc/kappa.php
- Seth N, Mahajan V, Kedia S, Sutar A, Sehgal K. Minimal Residual Disease (MRD) detection in B- ALL – Experience of a standalone flow cytometry laboratory. Pediatr Hematol Oncol. 2021;6(1):26–31. https://doi.org/10.1016/j.phoj.2020.12.003
- Singh J, Gorniak M, Grigoriadis G, et al. Correlation between a 10-color flow cytometric measurable residual disease (MRD) analysis and molecular MRD in adult B-acute lymphoblastic leukemia. Cytometry B Clin Cytom. 2022;102(2):115–122. https://doi.org/10.1002/cyto.b.22043
- Australasian Cytometry Society. ACS guideline for minimal residual disease testing [homepage on the Internet]. 2020 [cited 2025 Mar 30]. Available from: https://cytometry.org.au/wp-content/uploads/2020/08/ACS-MRD-GUIDELINES-2020-FINAL.pdf
- Van Dongen J, Van der Velden V, Brüggemann M, Orfao A. Minimal residual disease diagnostics in acute lymphoblastic leukemia: Need for sensitive, fast, and standardized technologies. Blood. 2015;125(26):3996–4009. https://doi.org/10.1182/blood-2015-03-580027
- Heuser M, Freeman S, Ossenkoppele G, et al. Update on MRD in acute myeloid leukemia: A consensus document from the European LeukemiaNet MRD working party. Blood. 2021;138(26):2753–2767. https://doi.org/10.1182/blood.2021013626
Appendix 1
| TABLE 1-A1: Details of cases that were true positive on flow cytometry. |
Appendix 2
| TABLE 1-A2: Details of cases that were false negative on flow cytometry. |
|