Original Research
Evaluation of residual disease detection by 10-colour flow cytometry using Navios, Navios EX and Kaluza C versus real-time quantitative polymerase chain reaction
Submitted: 22 January 2026 | Published: 13 October 2026
About the author(s)
Tsholofelo L. Malaka, Division of Molecular Medicine and Haematology, Faculty of Health Sciences, School of Pathology, University of the Witwatersrand, Johannesburg, South AfricaAnima Baiden, Division of Molecular Medicine and Haematology, Faculty of Health Sciences, School of Pathology, University of the Witwatersrand, Johannesburg, South Africa; and Department of Haematology, Charlotte Maxeke Johannesburg Academic Hospital, National Health Laboratory Services, Johannesburg, South Africa
Katherine E. Hodkinson, Division of Molecular Medicine and Haematology, Faculty of Health Sciences, School of Pathology, University of the Witwatersrand, Johannesburg, South Africa; Department of Haematology, Charlotte Maxeke Johannesburg Academic Hospital, National Health Laboratory Services, Johannesburg, South Africa; and Department of Haematology, Chris Hani Baragwanath Hospital, National Health Laboratory Services, Johannesburg, South Africa
Abstract
Background: Measurable residual disease (MRD) analysis is critical in acute lymphoblastic leukaemia (ALL) management protocols. Real-time quantitative polymerase chain reaction (RT-qPCR) is the accepted gold standard in MRD testing. Multi-parameter flow cytometry has emerged as a comparative testing modality. We evaluated the concordance of our centre’s 10-colour flow cytometry (Navios and Navios EX) with RT-qPCR (ABI 7500 Fast Real-Time PCR) for MRD detection.
Aim: To investigate the concordance of MRD detection using flow cytometry and RT-qPCR; and, separately, to compare immunoglobulin heavy chain (IgH) rearrangement PCR with RT-qPCR.
Setting: The National Health Laboratory Services Flow Cytometry and Somatic Cell Genetics Units at Charlotte Maxeke Johannesburg Academic Hospital.
Methods: Flow cytometry results between March 2020 and June 2023 were compared with RT-qPCR results in follow-up samples of patients with a diagnosis of B-cell acute lymphoblastic leukaemia (B-ALL), with one of the following genetic lesions: t(1;19), t(12;21) or t(9;22).
Results: A total of 74 bone marrow aspirate samples were eligible for inclusion. The flow cytometry assay showed a low sensitivity (13%) but a high specificity (98%) when compared with RT-qPCR. The positive and negative predictive values were 75% and 71% for flow cytometry versus qPCR, below the acceptable levels of 80% and 90%, respectively. Compared with RT-qPCR, the immunoglobulin heavy (IgH) assay demonstrated higher sensitivity (43%) and positive predictive value (91%), although sensitivity remained below levels for clinical utility.
Conclusion: The flow cytometry assay at our centre exhibited significantly lower sensitivity compared to PCR-based methods for the detection of MRD in follow-up cases. Among the PCR techniques, RT-qPCR demonstrated superior sensitivity for MRD detection in B-ALL.
Contribution: The research demonstrates the limitations of the current flow cytometry assay compared with RT-qPCR and identifies the need for improved MRD-tailored approaches, particularly in resource-limited settings where molecular testing may not be readily available.